nterstitial fluid glucose time-lag correction for real-time continuous glucose onitoring

@inproceedings{Keenana2012nterstitialFG,
  title={nterstitial fluid glucose time-lag correction for real-time continuous glucose onitoring},
  author={Barry Keenana and John J. Mastrototaroa and Stuart A. Weinzimerb and Garry M. Steil},
  year={2012}
}
Time lag between subcutaneous interstitial fluid and plasma glucose decreases the accuracy of realtime continuous glucose monitors. However, inverse filters can be designed to correct time lag and attenuate noise enabling the blood–glucose profile to be reconstructed in real time from continuous measurements of the interstitial-fluid glucose. We designed and tested a Wiener filter using a set of 20 sensor-glucose tracings (∼30 h each) with a 1-min sample interval. Delays of 10 ± 2 min (mean… 

Figures and Tables from this paper

Robust Nonlinear Model Predictive Control based on Constrained Saddle Point Optimization : Stability Analysis and Application to Type 1 Diabetes

This thesis deals with the design of a robust and safe control algorithm to aim at an artificial pancreas. More precisely we will be interested in controlling the stabilizing part of a classical

References

SHOWING 1-10 OF 32 REFERENCES

A novel approach to mitigating the physiological lag between blood and interstitial fluid glucose measurements.

The methodology presented here of using a pressure modulation technique to create an elevation in blood flow holds promise for significantly mitigating one of the most significant components of accuracy error for continuous monitoring systems.

Reconstruction of Glucose in Plasma from Interstitial Fluid Continuous Glucose Monitoring Data: Role of Sensor Calibration

Investigation of how well plasma glucose can be reconstructed from ISF CGM data found the quality of the reconstruction is unsatisfactory, and a recently proposed method to recalibrate ISF data was employed.

Graphical and numerical evaluation of continuous glucose sensing time lag.

Analysis of BG-CGM co-dynamics that is free from mathematical approximation of glucose fluctuations resulted in convenient visualization and numerical estimation of time lags typically associated with subcutaneous glucose sensing.

Modeling the Error of Continuous Glucose Monitoring Sensor Data: Critical Aspects Discussed through Simulation Studies

Modeling the statistical properties of CGM sensor errors from data collected in vivo is difficult because it requires perfect calibration and perfect knowledge of BG-to-IG dynamics, and this work shows that some assumptions made in the Breton and Kovatchev modeling approach may significantly affect the estimated sensor error and its statistical properties.

Interstitial fluid glucose dynamics during insulin-induced hypoglycaemia

The elevated ratio suggests that the plasma to ISF glucose gradient was decreased following hypoglycaemia, possibly due to increased skin blood flow.

Determination of plasma glucose during rapid glucose excursions with a subcutaneous glucose sensor.

The subcutaneous glucose sensor has the potential to facilitate the detection of hypoglycemia and improve overall glycemic control when used in a real-time monitor and the rapid response should be sufficient to allow a fully automated closed-loop insulin delivery system to be developed based on the sub cutaneous sensing site.

Enhanced accuracy of continuous glucose monitoring by online extended kalman filtering.

Improved Bayesian calibration method takes into account the existence of BG-to-IG kinetics by incorporating a population convolution model and exploits only four BG reference samples per day and significantly improves the accuracy of CGM signals.

Subcutaneous glucose predicts plasma glucose independent of insulin: implications for continuous monitoring.

It is concluded that the relationship between blood and ISF glucose is not affected by insulin and that delays inISF glucose equilibration can be corrected with digital filters.

Can interstitial glucose assessment replace blood glucose measurements?

The differences between plasma and ISF glucose will not be a significant obstacle in advancing the use of ISF as an alternative to blood glucose measurements, according to a model-based approach.