Volume blood flow measurements with a transit time flowmeter: an in vivo and in vitro variability and validation study.

@article{Lundell1993VolumeBF,
  title={Volume blood flow measurements with a transit time flowmeter: an in vivo and in vitro variability and validation study.},
  author={Anders Lorentz Lundell and David Bergqvist and Eva Mattsson and Bo N Nilsson},
  journal={Clinical physiology},
  year={1993},
  volume={13 5},
  pages={547-57}
}
A transit time flowmeter, Transonic TC101 (Transonic Inc., USA) has been evaluated to study volume flow. Two flow probes, 4 and 6 mm, were used. The experiments were carried out in vivo on the carotid arteries (n = 16) in sheep and in vitro against a pre calibrated rotational pump. Exsanguination measurements, n = 32, were used for the calculation of the error of measurement in vivo. The error of measurement in vitro was calculated on 11 different flow rates from 0 to 150 ml min-1. The… CONTINUE READING