Verification and quality control of routine hematology analyzers
@article{Vis2016VerificationAQ,
title={Verification and quality control of routine hematology analyzers},
author={Jolande Y. Vis and Albert Huisman},
journal={International Journal of Laboratory Hematology},
year={2016},
volume={38}
}Verification of hematology analyzers (automated blood cell counters) is mandatory before new hematology analyzers may be used in routine clinical care. The verification process consists of several items which comprise among others: precision, accuracy, comparability, carryover, background and linearity throughout the expected range of results. Yet, which standard should be met or which verification limit be used is at the discretion of the laboratory specialist. This paper offers practical…
Topics from this paper
58 Citations
Evaluation of the performance of sysmex XN-3100 automated hematology analyzer regarding the sysmex XE-2100 and microscopic examination
- Medicine
- 2020
The Sysmex XN-3100 not only showed a strong correlation and agreement with the XE-2100 but also displayed a comparable analytical sensitivity, and increased specificity, which may result in an improved turnaround time and throughpu.
Multicenter performance evaluation of the Abbott Alinity hq hematology analyzer
- MedicineClinical chemistry and laboratory medicine
- 2019
Analytical performance of Alinity hq was shown to be consistent with state-of-the-art (SOTA) performance characteristics, and good overall agreement with results obtained with CELL-DYN Sapphire, as well as manual WBC differential.
Uncertenity Evaluation of boold analyser system
- Medicine2020 International Conference on Computer, Control, Electrical, and Electronics Engineering (ICCCEEE)
- 2021
The methodology processed in this research studies the chemical and physical characteristics, and records the effect of the reagents on the sample, and the accuracy of hematology device measurement.
Evaluation of humoral pattern detection performance of Mindray BC-6000PLUS blood analyzer based on WS/T 662-2020.
- MedicineAnnals of palliative medicine
- 2021
All performance indicators of the Mindray BC-6000PLUS automatic blood analyzer met the requirements of the manufacturer's criteria and can fulfill the requirement of body fluid sample routine test in clinical practice.
Policies and practices in the field of laboratory hematology in Croatia – a current overview and call for improvement
- MedicineClinical chemistry and laboratory medicine
- 2021
The survey revealed substantial differences in all aspects of laboratory hematology practices among Croatian medical laboratories, indicating the need for universal recommendations at the national level.
Evaluation of automated hematology analyzer DYMIND DH76 compared to SYSMEX XN 1000 system
- MedicineJournal of medical biochemistry
- 2021
With excellent analytical performance and acceptable comparative analysis, DYMIND DH76 hematology analyser covered the predefined international standards and requirements and is fully appropriate for clinical application.
Performance evaluation of the new measurement channels on the automated Sysmex XN-9000 hematology analyzer.
- MedicineClinical biochemistry
- 2018
Preliminary evaluation of a new flow cytometry method for the routine hematology workflow
- MedicineClinical chemistry and laboratory medicine
- 2019
The new hybrid method generates a leukocytes differential count suitable for routine clinical use and it is also useful for identifying morphological abnormalities with a reduction in cost and improvement of screening for first level hematology workflow.
The use of routine hematology analyzers for quality management of leukocyte depletion in blood transfusion products: a comment
- Biology
- 2020
The authors of the discussed paper have investigated whether a hematology analyzer is suitable for monitoring the leukocyte depletion process in the preparation of blood products for human transfusion and have demonstrated that the type of hematological analyzer discussed in their paper is suitable.
Critical Systematic Errors in the Implementation and Follow-up of Performance in the Hematology Area: A Prospective Study
- MedicineRevista Medicina Universitaria
- 2019
The quality evaluation of the ΔSE analysis showed a heterogeneous performance and intra-parameter control levels, as well as violations of the Westgard rules and mismatches in the determination of means (calculated and automatic).
References
SHOWING 1-10 OF 47 REFERENCES
Verification and standardization of blood cell counters for routine clinical laboratory tests.
- MedicineClinics in laboratory medicine
- 2015
Quality specification in haematology: the automated blood cell count.
- MedicineClinica chimica acta; international journal of clinical chemistry
- 2004
Evaluation of the fully automated hematological analyzer Mindray BC 6800: comparison with Horiba ABX Pentra DX120
- MedicineInternational journal of laboratory hematology
- 2014
The aim of this study was to evaluate the performance of BC-6800 hematology analyzer, a new generation fully automated hematological analyzer in use in this laboratory, and in comparison with ABX Pentra DX120, a reference fully automated Hematology Analyzer used in this lab.
Comparison of five automated hematology analyzers in a university hospital setting: Abbott Cell-Dyn Sapphire, Beckman Coulter DxH 800, Siemens Advia 2120i, Sysmex XE-5000, and Sysmex XN-2000
- MedicineClinical chemistry and laboratory medicine
- 2015
This is the most comprehensive side-by-side comparison of five current top of the range routine hematology analyzers in the setting of a university hospital central laboratory and shows a good concordance for basic blood count parameters.
Evaluation of the Abbott CELL-DYN 4000 hematology analyzer.
- MedicineAmerican journal of clinical pathology
- 2000
The evaluation of leukocyte differential counts indicated an excellent correlation with the manual reference method for neutrophils and lymphocytes, a good correlation for monocytes and eosinophils, and a poor correlation for basophils in samples with low counts, but an improvement of the correlation coefficient was found.
ICSH guidelines for the evaluation of blood cell analysers including those used for differential leucocyte and reticulocyte counting
- MedicineInternational journal of laboratory hematology
- 2014
This revision is intended to update the 1994 ICSH guidelines but is updated to include new methods, such as digital image analysis for blood cells, a flow cytometric method intended to replace the reference manual 400 cell differential, and numerous new cell indices not identified morphologically are introduced.
Automated blood cell counts: state of the art.
- MedicineAmerican journal of clinical pathology
- 2008
The state of the art regarding traditional and new parameters with emphasis on clinical applications and analytic quality are analyzed regarding the CBC count and leukocyte differential count.
Hematology analyzers: special considerations for pediatric patients.
- MedicineClinics in laboratory medicine
- 2015
Performance evaluation of the Sysmex haematology XN modular system
- MedicineJournal of Clinical Pathology
- 2012
The Sysmex XN haematology instrument showed reduced sample turnaround time of 10% and reduced number of blood films for examination, 49% less than the XE-2100 without loss of sensitivity with more precise and accurate results on low cell counts.