          F175-8
        }
}
This study validates the accuracy of miniature ultrasonic transit-time flow probes for measuring renal blood flow (RBF) in the rat. Probes for 1-mm and 2-mm vessels were calibrated ex vivo using excised arteries at varying flow rates and hematocrit (Hct). Correlation between measured and true flow rates for the 2-mm probe were identical (r = 1.0) at both normal and subnormal Hct values. Correlation for the 1-mm probe was high (r = 0.994) at normal Hct, but varied at both high flow rates and… CONTINUE READING