Validation of a novel particle isolation procedure using particle doped tissue samples

@inproceedings{Patel2018ValidationOA,
  title={Validation of a novel particle isolation procedure using particle doped tissue samples},
  author={Jatan Patel and Samridhi Lal and S P Wilshaw and R M Hall and J. L. Tipper},
  booktitle={Data in brief},
  year={2018}
}
A novel particle isolation method for tissue samples was developed and tested using particle-doped peri-articular tissues from ovine cadavers. This enabled sensitivity of the isolation technique to be established by doping tissue samples of 0.25 g with very low particle volumes of 2.5 µm3 per sample. Image analysis was used to verify that the method caused no changes to particle size or morphologies. 

From This Paper

Figures, tables, and topics from this paper.