Using gas modifiers to significantly improve sensitivity and selectivity in a cylindrical FAIMS device.

@article{Purves2014UsingGM,
  title={Using gas modifiers to significantly improve sensitivity and selectivity in a cylindrical FAIMS device.},
  author={Randy W. Purves and Allison R Ozog and Stephen J. Ambrose and Satendra Prasad and Michael W Belford and J. C. Dunyach},
  journal={Journal of the American Society for Mass Spectrometry},
  year={2014},
  volume={25 7},
  pages={
          1274-84
        }
}
Recent reports describing enhanced performance when using gas additives in a DMS device (planar electrodes) have indicated that comparable benefits are not attainable using FAIMS (cylindrical electrodes), owing to the non-homogeneous electric fields within the analyzer region. In this study, a FAIMS system (having cylindrical electrodes) was modified to allow for controlled delivery of gas additives. An experiment was carried out that illustrates the important distinction between gas modifiers… CONTINUE READING