Uniformity of response of Uranium fission chambers used as neutron beam monitors
@article{Lafont2021UniformityOR, title={Uniformity of response of Uranium fission chambers used as neutron beam monitors}, author={Fabien Lafont and B. von Gu{\'e}rard and Richard J. Hall-Wilton and Kalliopi Kanaki}, journal={Journal of Instrumentation}, year={2021}, volume={17} }
Uranium fission chambers are commonly used for neutron beam monitoring in neutron research facilities. The challenges brought with the renewing or commissioning of new neutron facilities encourage to minimize the interaction between the monitor and the neutron beam. In order to characterize both their detection efficiency and transmission factor, several fission chambers, have been scanned on the CT1 monochromatic beam line of the ILL. Results show that the monitor transparency, measured at the…
One Citation
Multitube monitors: a new-generation of neutron beam monitors
- PhysicsJournal of Instrumentation
- 2022
With the renewal of many neutron science instruments and the commissioning of new neutron facilities, there is a rising demand for improved neutron beam monitoring systems with reduced beam…
References
SHOWING 1-2 OF 2 REFERENCES