Translational and rotational intra- and inter-fractional errors in patient and target position during a short course of frameless stereotactic body radiotherapy.

@article{Josipovic2012TranslationalAR,
  title={Translational and rotational intra- and inter-fractional errors in patient and target position during a short course of frameless stereotactic body radiotherapy.},
  author={Mirjana Josipovic and Gitte Persson and Ashildur Logadottir and Bob Smulders and Gunnar Westmann and Jens Peter Bangsgaard},
  journal={Acta oncologica},
  year={2012},
  volume={51 5},
  pages={
          610-7
        }
}
BACKGROUND Implementation of cone beam computed tomography (CBCT) in frameless stereotactic body radiotherapy (SBRT) of lung tumours enables setup correction based on tumour position. The aim of this study was to compare setup accuracy with daily soft tissue matching to bony anatomy matching and evaluate intra- and inter-fractional translational and rotational errors in patient and target positions. MATERIAL AND METHODS Fifteen consecutive SBRT patients were included in the study. Vacuum… CONTINUE READING