Three‐dimensional dosimetry of TomoTherapy by MRI‐based polymer gel technique

@inproceedings{Watanabe2010ThreedimensionalDO,
  title={Three‐dimensional dosimetry of TomoTherapy by MRI‐based polymer gel technique},
  author={Yoichi Watanabe and Natanasabapathi Gopishankar},
  booktitle={Journal of applied clinical medical physics},
  year={2010}
}
Verification of the dose calculation model and the software used for treatment planning is an important step for accurate radiation delivery in radiation therapy. Using BANG3 polymer gel dosimeter with a 3 Tesla magnetic resonance imaging (MRI) scanner, we examined the accuracy of TomoTherapy treatment planning and radiation delivery. We evaluated one prostate treatment case and found the calculated three-dimensional (3D) dose distributions agree with the measured 3D dose distributions with an… CONTINUE READING

From This Paper

Figures, tables, and topics from this paper.

Citations

Publications citing this paper.
Showing 1-9 of 9 extracted citations

Uncertainty Analysis in MRI-based Polymer Gel Dosimetry

Journal of biomedical physics & engineering • 2017
View 4 Excerpts
Highly Influenced

References

Publications referenced by this paper.
Showing 1-10 of 13 references

Comparison of R2 estimation algorithms for polymer gel dosimetry [abstract

Y Watanabe, H. Kubo
Med Phys • 2009
View 1 Excerpt

History of tomotherapy.

Physics in medicine and biology • 2006
View 1 Excerpt