The value of routine real ear measurement of the gain of digital hearing aids.

@article{Aazh2007TheVO,
  title={The value of routine real ear measurement of the gain of digital hearing aids.},
  author={Hashir Aazh and Brian C. J. Moore},
  journal={Journal of the American Academy of Audiology},
  year={2007},
  volume={18 8},
  pages={
          653-64
        }
}
  • Hashir AazhB. Moore
  • Published 1 September 2007
  • Medicine
  • Journal of the American Academy of Audiology
The main aims of this study were: (1) to determine whether routine real ear insertion gain (REIG) measurement is necessary in fitting digital hearing aids; and (2) to assess the extent to which modifying the frequency-gain response of an aid can lead to better matches to the target in cases where the target gain was not initially achieved. The target formula was selected as NAL-NL1 in the programming software of four types of digital hearing aids. REIG measurements on 42 ears showed that 64% of… 

Figures and Tables from this paper

The accuracy of matching target insertion gains with open-fit hearing aids.

It is suggested that target insertion gains for the open-fit hearing aids used here are rarely achieved with a first fitting but can usually be achieved through adjustments based on REIG measurements.

Comparison of Personal Sound Amplification Products to Professionally-Fitted Hearing Aids

This study was the first of its kind to compare a self-fitted personal sound amplification product (PSAP) with professionally-fitted hearing aids in terms of real-ear aided gain, audibility, and speech understanding in background noise.

Verification of Hearing Aids by Comparing Real Ear Measurements and Word Recognition Scores

This study verified HAs that have gone through the fitting program using speech mapping REM and compared the outcome with word recognition scores (WRS) to evaluate functional gain.

Achieved Gain and Subjective Outcomes for a Wide-Bandwidth Contact Hearing Aid Fitted Using CAM2

Most participants preferred slightly less gain at 6 to 10 kHz than prescribed for experienced users by CAM2, preferring similar gains to those prescribed for inexperienced users, but gains over the extended high frequencies were high relative to those that are currently available with acoustic hearing aids.

Comparison: real and simulated ear insertion gain

HA software programming simulated insertion gain was higher than real ear insertion gain obtained with probe microphone measures, and the use of real ear measures is important during verification of HA.

Using probe-microphone measurements to improve the match to target gain and frequency response slope, as a function of earmould style, frequency, and input level

The results indicate that IF settings are inadequate, at least for the model of hearing aid used in the present study; however, significant discrepancies remained, even after adjustment.

Discrimination of Gain Increments in Speech-Shaped Noises

To examine what gain adjustments are discriminable, the just-noticeable differences (JNDs) for level increments in speech-shaped noises processed with prescription gains are measured and suggest noticeable fine-tuning adjustments of 3 dB and self-fitting adjustments of 5‬dB.

Modeling the Influence of Acoustic Coupling of Hearing Aids on FM Signal-to-Noise Ratio.

A model was developed to examine variables that influence signal-to-noise ratio (SNR) at the tympanic membrane when using a hearing aid (HA) and frequency modulated (FM) system and predicts similar overall SNRs at the TM regardless of coupling method when HA and FM microphones are active.

Addressing age related hearing loss through engineering accessible and affordable hearing technology

The proposed design addresses a long-standing and grand challenge of affordable and accessible hearing technology for every elderly person on this planet with the LoCHAid, designed to be an over-the-counter self-serviceable solution for elderly individuals with ARHL.

Clinical Verification of Hearing Aid Performance

There is an urgent clinical need to address the wide range of clinical outcomes that occur in hearing aid users with apparently similar characteristics and a degree of personalization to the target values beyond that associated with hearing threshold levels.
...

References

SHOWING 1-10 OF 34 REFERENCES

The value of routine in-the-ear measurement of hearing aid gain.

Though there were audiometric differences between those who achieved satisfactory gain and those who did not, there was too much overlap between groups to make any audiometric index or combination of indices of value in predicting the likelihood of failure to achieve target gain.

Hearing aid software predictive gain values: How accurate are they?

There are several ways to assess the electroacoustic characteristics of hearing aids, but none of these on-screen representations have been made with the actual hearing aid; rather, they represent simulations based on average and expected performance values.

Individualized correction factors in the preselection of hearing aids.

This study investigated the extent to which inconsistencies in the sound-field reference position can affect comparative corrections for the real-ear unaided response (REUR) and the adequacy of using KEMAR estimates of the effects of the location of the hearing aid microphone.

Evaluation of the National Acoustic Laboratories' new hearing aid selection procedure.

The formula was found to be highly effective, but there were some cases where a change in aid prescription was indicated, and a simple evaluation procedure using paired-comparison judgments is proposed for detecting such cases.

The National Acoustic Laboratories' (NAL) New Procedure for Selecting the Gain and Frequency Response of a Hearing Aid

It is concluded that the new formula for selecting the gain and frequency response of a hearing aid should prescribe a near optimal frequency response with few exceptions.

Comparison of three procedures for initial fitting of compression hearing aids. I. Experienced users, fitted bilaterally

The effectiveness of three procedures for the initial fitting of hearing aids with multiband compression are compared, concluding that the CAMEQ and CAMREST procedures provide more appropriate initial fittings than DSL I/O.

Predicting hearing aid response in real ears.

  • P. E. Sanborn
  • Physics
    The Journal of the Acoustical Society of America
  • 1998
The theoretical description of an electrical analog model of the ear and hearing aid system is given and it is likely to be valid from low frequencies to at least 10 kHz.

Is probe‐mic measurement of HA gain‐frequency response best practice?

There are arguments to be made both for and against the routine measurement of real-ear gain and/or output in hearing aid fitting.

Insertion Gain Measurements using Two Low-Powered Analogue Hearing Aids

Abstract Differences between insertion gain measurements and coupler measurements were investigated using two different hearing aids, to assess the accuracy of hearing aid fitting methods which

Use of a loudness model for hearing-aid fitting. I. Linear hearing aids.

A model for predicting loudness for people with cochlear hearing loss is applied to the problem of prescribing the frequency-gain characteristic of a linear hearing aid and a new formula - the 'Cambridge formula' - is developed for prescribing insertion gain from audiometric thresholds.