The simultaneous measurement of low rates of CO2 and O2 exchange in biological systems.

@article{Willms1997TheSM,
  title={The simultaneous measurement of low rates of CO2 and O2 exchange in biological systems.},
  author={J. Willms and Andrea Dowling and Zhao Ming Dong and Steven T Hunt and Barry J. Shelp and David Layzell},
  journal={Analytical biochemistry},
  year={1997},
  volume={254 2},
  pages={272-82}
}
An instrument for measuring low rates of biological O2 exchange using an open-flow gas analysis system is described. A novel differential O2 sensor that is capable of measuring as little as 0.4 Pa O2 against a back-ground of ambient air (20,900 Pa O2), yet has a dynamic range of +/- 2000 Pa O2 (i.e., +/- ca. 2% O2) is described. Baseline drift was typically less than 0.025 Pa min-1. The differential O2 sensor was incorporated into a respiratory quotient/photosynthetic quotient analyzer that… CONTINUE READING