The peripheral dose outside the applicator in electron beams of Oncor linear accelerator.

@article{Iktueren2012ThePD,
  title={The peripheral dose outside the applicator in electron beams of Oncor linear accelerator.},
  author={Basak Iktueren and Hatice Bilge and Songul Karacam and Gulyuz Atkovar},
  journal={Radiation protection dosimetry},
  year={2012},
  volume={150 2},
  pages={
          192-7
        }
}
In this study, the peripheral dose outside the applicator was measured using electron beams produced by an Oncor linear accelerator and compared with the data of the treatment planning system (TPS). The dose profiles have been measured, by using a water-equivalent slab phantom and a parallel plate ionisation chamber, at 6, 9 and 15 MeV energy levels in 5×5, 10×10, 15×15, 20×20 and 25×25 cm(2) applicators and at 0, 10 and 20° gantry angles; and at the surface, 0.2, 0.5, 1 cm and d(max) depth for… CONTINUE READING