The optimum size of prosthetic heads in relation to the wear of plastic sockets in total replacement of the hip

@article{Charnley1969TheOS,
  title={The optimum size of prosthetic heads in relation to the wear of plastic sockets in total replacement of the hip},
  author={John Charnley and A. Kamangar and M. D. Longfield},
  journal={Medical and biological engineering},
  year={1969},
  volume={7},
  pages={31-39}
}
A study has been made of the wear of hip sockets of P.T.F.E. against various sizes of stainless steel femoral head in the course of total prosthetic replacement of the hip in the human subject. The most serious aspect of the failure of P.T.F.E. was the production of a tissue reaction by wear debris. Artificial hip joints using plastic sockets should, therefore, be designed to produce the minimum volume of wear debris over a given period. The smaller the diameter of the metal head the smaller… CONTINUE READING