The new haematology analyzer DxH 800: an evaluation of the analytical performances and leucocyte flags, comparison with the LH 755.

@article{Jean2011TheNH,
  title={The new haematology analyzer DxH 800: an evaluation of the analytical performances and leucocyte flags, comparison with the LH 755.},
  author={Arnaud Jean and Catherine Boutet and Bernard Lenormand and M-P Callat and G{\'e}rard Buchonnet and Virginie Barbay and J-P Basuyau and Michael Vasse},
  journal={International journal of laboratory hematology},
  year={2011},
  volume={33 2},
  pages={
          138-45
        }
}
INTRODUCTION The analytical performance and the abnormality messages on differential (flags) of the new analyzer Beckman Coulter DxH 800 were compared with those of the LH 755. METHODS First, we evaluated the accuracy of the results of the DxH 800, in comparison with the LH 755, in 125 samples without alarm using unflagged sample results on both analyzers. Second, flagged samples on the LH 755 but not flagged by the DxH 800 were evaluated by flow cytometry for accuracy of the DxH 800 results… CONTINUE READING