The comparison of radiation dose between C-arm flat-detector CT (DynaCT) and multi-slice CT (MSCT): a phantom study.

@article{Bai2012TheCO,
  title={The comparison of radiation dose between C-arm flat-detector CT (DynaCT) and multi-slice CT (MSCT): a phantom study.},
  author={M Pavithra Bai and Bin Liu and Hongqiu Mu and Xianghua Liu and Yuan Jiang},
  journal={European journal of radiology},
  year={2012},
  volume={81 11},
  pages={3577-80}
}
PURPOSE To assess and compare the radiation dose of DynaCT in phantoms to that of MSCT. METHODS AND MATERIALS A male Anderson Radiation Therapy 200 phantom with embedded thermoluminescence dosimeters (TLDs) was scanned with DynaCT (Axiom Artis dTA, Siemens Healthcare, Forchheim, Germany) and MSCTs (MSCT1: Lightspeed VCT, GE, Milwaukee, USA; MSCT2: Sensation Cardiac 64, Siemens Healthcare, Erlangen, Germany), respectively. For DynaCT, radiation exposure data with automatic exposure control… CONTINUE READING