Test-retest reliability of the MicroFET 4 hand-grip dynamometer.

@article{Bohannon2006TestretestRO,
  title={Test-retest reliability of the MicroFET 4 hand-grip dynamometer.},
  author={Richard W. Bohannon},
  journal={Physiotherapy theory and practice},
  year={2006},
  volume={22 4},
  pages={219-21}
}
This study was conducted to determine the test-retest reliability of measurements of grip strength obtained with the MicroFET 4 dynamometer. Measurements did not differ significantly between days and had high intraclass correlation coefficients (>.970). The technical error of measurement was low (< or =22.7 newtons). On the basis of results from this convenience sample of nondisabled adults, it would appear that measurements obtained with the MicroFET 4 possess high test-retest reliability.