Stiffness and endurance of a locking compression plate fixed on fractured femur.

@article{Kanchanomai2010StiffnessAE,
  title={Stiffness and endurance of a locking compression plate fixed on fractured femur.},
  author={Chaosuan Kanchanomai and Panurungsit Muanjan and Vajara Phiphobmongkol},
  journal={Journal of applied biomechanics},
  year={2010},
  volume={26 1},
  pages={
          10-6
        }
}
The effects of locking screw position (long column fixation--long distance between the nearest screws to the fracture--and short column fixation--short distance between the nearest screws to the fracture) and fracture gap size (1-mm and 8-mm transverse fracture gap) on stiffness and fatigue of fractured femur fixed with a locking compression plate (LCP) were biomechanically evaluated. The stiffness of 1-mm fracture gap models and that of intact femoral model were in the range of 270-284 N/mm… CONTINUE READING

Figures, Tables, and Topics from this paper.

References

Publications referenced by this paper.
SHOWING 1-10 OF 16 REFERENCES

Failure analysis of a commercially pure titanium plate for osteosynthesis

C.R.F. Azevedo
  • Engineering Failure Analysis,
  • 2003
VIEW 2 EXCERPTS

Similar Papers