Specification of acoustic output level and measurement uncertainty in ultrasonic exposimetry

@article{Ziskin2003SpecificationOA,
  title={Specification of acoustic output level and measurement uncertainty in ultrasonic exposimetry},
  author={M C Ziskin},
  journal={IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control},
  year={2003},
  volume={50},
  pages={1023-1034}
}
  • M C Ziskin
  • Published 2003 in
    IEEE Transactions on Ultrasonics, Ferroelectrics…
This manuscript was written to bring together the pertinent statistical procedures for specifying acoustic output values and measurement uncertainties of ultrasound devices, and to present them in a way that would be clear and convenient to the reader. Implementations of these procedures are illustrated in several examples: measurement of ultrasonic power of a single transducer, measurement of the power output of a transducer model, and measurement of the power output of an ultrasound scanner… CONTINUE READING
6 Citations
8 References
Similar Papers

References

Publications referenced by this paper.
Showing 1-8 of 8 references

and F

  • S. E. Fick, D. G. Eitzen, C. E. Tschiegg
  • R. Breckenbridge, “An ultrasonic absolute power…
  • 1984
Highly Influential
8 Excerpts

Measurement uncertainty in ultrasonic exposimetry,

  • M. C. Ziskin
  • Boca Raton, FL: CRC Press,
  • 1993
1 Excerpt

Statistical Principles in Experimental Design

  • B. J. Winer, D. B. Brown, K. M. Michels
  • 3rd ed. New York: McGraw-Hill
  • 1991
1 Excerpt

and R

  • R. C. Preston, D. R. Bacon
  • A. Smith, “Calibration of medical ultrasonic…
  • 1988
1 Excerpt

Statistical Methods

  • G. W. Snedecor, W. G. Cochran
  • 7th ed. Ames, IA: Iowa State Univ. Press
  • 1980

Experimental Statistics

  • M. G. Natrella
  • Handbook 91. Washington, D.C.: National Bureau of…
  • 1963
3 Excerpts

Similar Papers

Loading similar papers…