Spatial spread of neural excitation in cochlear implant recipients: comparison of improved ECAP method and psychophysical forward masking

@article{Cohen2003SpatialSO,
  title={Spatial spread of neural excitation in cochlear implant recipients: comparison of improved ECAP method and psychophysical forward masking},
  author={Lawrence T. Cohen and Louise M Richardson and Elaine Saunders and Robert Cowan},
  journal={Hearing Research},
  year={2003},
  volume={179},
  pages={72-87}
}
This study introduces and evaluates a method for measurement of the longitudinal spread of electrically evoked neural excitation in the cochlea, using the Neural Response Telemetry system (NRT) available with the Nucleus((R)) 24 cochlear implant system. The recently released version of the NRT software (version 3.0) enables presentation of the 'masker' and 'probe' on different electrodes. In the present method the probe position was fixed, while the masker position was varied across the… CONTINUE READING
Highly Cited
This paper has 127 citations. REVIEW CITATIONS

From This Paper

Topics from this paper.

Citations

Publications citing this paper.
Showing 1-10 of 60 extracted citations

Modelling encapsulation tissue around cochlear implant electrodes

Medical and Biological Engineering and Computing • 2005
View 3 Excerpts
Highly Influenced

127 Citations

051015'06'09'12'15'18
Citations per Year
Semantic Scholar estimates that this publication has 127 citations based on the available data.