Sparse 2-D arrays for 3-D phased array imaging - experimental validation

@article{Austeng2002Sparse2A,
  title={Sparse 2-D arrays for 3-D phased array imaging - experimental validation},
  author={Andreas Austeng and Sverre Holm},
  journal={IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control},
  year={2002},
  volume={49},
  pages={1087-1093}
}
To be able to describe more precisely the behavior of a real-time 3-D ultrasound system with either a dense array or various sparse designs, experimental data from a 2-D fully connected array prototype with 50/spl times/50 elements have been collected. The data have been processed off line to form synthetic aperture 3-D volume images. Simulated and experimental results are compared and show good correlation. The performance of the best sparse designs, all thinned to more than 50%, offer… CONTINUE READING

From This Paper

Figures, tables, and topics from this paper.

Citations

Publications citing this paper.
Showing 1-10 of 12 extracted citations

References

Publications referenced by this paper.
Showing 1-10 of 30 references

A two-dimensional transducer array for real-time 3D medical ultrasound imaging

  • J.-M. Bureau, W. Steichen, G. Lepail
  • Proc. IEEE Ultrason. Symp., pp. 1065–1068, 1998.
  • 1998
Highly Influential
4 Excerpts

New opportunities in ultrasonic transducers emerging from innovations in piezoelectric materials

  • W. A. Smith
  • SPIE, F. L. Lizzi, Ed. vol. 1733, pp. 3–16, Nov…
  • 1992
Highly Influential
5 Excerpts

2-D arrays

  • O. T. von Ramm
  • Proc. Ultrason. Int. 1999 Joint 1999 World Congr…
  • 1999
1 Excerpt

A spiral 2D phased array for 3D imaging

  • T. S. Sumanaweera, J. Schwartz, D. Napolitano
  • Proc. IEEE Ultrason. Symp., vol. 2, pp. 1271–1274…
  • 1999
1 Excerpt

Update of two dimensional arrays for real time volumetric and real time intracardiac imaging

  • E. D. Light, J. O. Fiering, P. A. Hultman, W. Lee, S. W. Smith
  • Proc. IEEE Ultrason. Symp., pp. 1217–1220, Oct…
  • 1999
1 Excerpt

A two - dimensional transducer array for real - time 3 D medical ultrasound imag

  • W. Steichen J.-M. Bureau, G. Lepail
  • Proc . IEEE Ultrason . Symp .
  • 1998

Similar Papers

Loading similar papers…