Some characteristics of a new multiple-channel photometer for through-the-plate reading of microplates to be used in enzyme-linked immunosorbent assay.

Abstract

The performance of a prototype of a now commercially available multiple-channel photometer for high-speed reading of microplates to be used in the enzyme-linked immunosorbent assay was evaluated. For this purpose the extinction of a stable test liquid dispensed in flat-bottomed polystyrene microplates was measured, varying the test volume and the optical density. From the results obtained for dilutions of a single, biolgoically inactive material we concluded that the photometric inaccuracy is negligible when performing enzyme-linked immunosorbent assay with biologically active samples.

Cite this paper

@article{Ruitenberg1980SomeCO, title={Some characteristics of a new multiple-channel photometer for through-the-plate reading of microplates to be used in enzyme-linked immunosorbent assay.}, author={E. Joost Ruitenberg and V M Sekhuis and Berry J Brosi}, journal={Journal of clinical microbiology}, year={1980}, volume={11 2}, pages={132-4} }