Signal to noise ratio and uncertainty in diffusion tensor imaging at 1.5, 3.0, and 7.0 Tesla.

@article{Polders2011SignalTN,
  title={Signal to noise ratio and uncertainty in diffusion tensor imaging at 1.5, 3.0, and 7.0 Tesla.},
  author={Daniel Louis Polders and Alexander Leemans and Jeroen Hendrikse and Manus J. Donahue and Peter R. Luijten and Johannes M. Hoogduin},
  journal={Journal of magnetic resonance imaging : JMRI},
  year={2011},
  volume={33 6},
  pages={1456-63}
}
PURPOSE To compare diffusion tensor imaging (DTI) measurements at ultra high field strength (7 Tesla [T]) in human volunteers with DTI measurements performed at 1.5 and 3 Tesla. MATERIALS AND METHODS The signal to noise ratio (SNR) and the uncertainty in fitted DTI parameters fractional anisotropy and primary eigenvector are assessed with tractography based regions of interest, measured in nine volunteers at 1.5T, 3T, and 7T with clinically available hardware configurations. RESULTS An… CONTINUE READING