Separation of cisplatin and its hydrolysis products using electrospray ionization high-field asymmetric waveform ion mobility spectrometry coupled with ion trap mass spectrometry.

@article{Cui2003SeparationOC,
  title={Separation of cisplatin and its hydrolysis products using electrospray ionization high-field asymmetric waveform ion mobility spectrometry coupled with ion trap mass spectrometry.},
  author={Meng Cui and Luyi Ding and Zolt{\'a}n Mester},
  journal={Analytical chemistry},
  year={2003},
  volume={75 21},
  pages={5847-53}
}
Cisplatin and its mono- and dihydrated complexes have been separated using a high-field asymmetric waveform ion mobility spectrometry (FAIMS) analyzer interfaced with electrospray ionization (ESI) and ion trap mass spectrometry (ITMS). The addition of helium to the nitrogen curtain/carrier gas in the FAIMS device improved both the sensitivity and… CONTINUE READING