Sensitivity and specificity of blood cobalt and chromium metal ions for predicting failure of metal-on-metal hip replacement.

@article{Hart2011SensitivityAS,
  title={Sensitivity and specificity of blood cobalt and chromium metal ions for predicting failure of metal-on-metal hip replacement.},
  author={Alister Hart and Shiraz A. Sabah and Ashwath S. Bandi and Paolo Maggiore and Payam Tarassoli and Barry A. Sampson and J A Skinner},
  journal={The Journal of bone and joint surgery. British volume},
  year={2011},
  volume={93 10},
  pages={
          1308-13
        }
}
Blood metal ions have been widely used to investigate metal-on-metal hip replacements, but their ability to discriminate between well-functioning and failed hips is not known. The Medicines and Healthcare products Regulatory Agency (MHRA) has suggested a cut-off level of 7 parts per billion (ppb). We performed a pair-matched, case-control study to investigate the sensitivity and specificity of blood metal ion levels for diagnosing failure in 176 patients with a unilateral metal-on-metal hip… CONTINUE READING

Citations

Publications citing this paper.
SHOWING 1-10 OF 79 CITATIONS

The safety of Metal-on-Metal joint replacements with a particular focus on hip implants

  • 2014
VIEW 5 EXCERPTS
CITES BACKGROUND & METHODS
HIGHLY INFLUENCED

Pleomorphic clinical spectrum of metallosis in total hip arthroplasty

Francesco Pisanu, Carlo Doria, +4 authors Paolo Tranquilli Leali
  • International Orthopaedics
  • 2018
VIEW 2 EXCERPTS
CITES BACKGROUND

FILTER CITATIONS BY YEAR

2012
2019

CITATION STATISTICS

  • 3 Highly Influenced Citations

  • Averaged 10 Citations per year over the last 3 years

References

Publications referenced by this paper.
SHOWING 1-10 OF 25 REFERENCES

5th annual report

No authors listed. National Joint Registry for England, Wales
  • 2008:78. www.njrcentre.org.uk
  • 2011

all metal-on-metal (MOM) hip replacements, 2010 MDA/2010/ 033

No authors listed. Medicines, Healthcare products Regulatory Agency . Medical Device Alert
  • http://www.mhra.gov.uk/Publications/Safetywarnings/MedicalDeviceAlerts/ CON079157
  • 2011

The painful metal-on-metal hip resurfacing.

  • The Journal of bone and joint surgery. British volume
  • 2009