Semi-automated vs. visual reading of urinalysis dipsticks.

@article{Peele1977SemiautomatedVV,
  title={Semi-automated vs. visual reading of urinalysis dipsticks.},
  author={J D Peele and Richard H. Gadsden and R H Crews},
  journal={Clinical chemistry},
  year={1977},
  volume={23 12},
  pages={2242-6}
}
Semi-automated reflectance readings of urinalysis dipsticks showed better precision than visual readings in the case of artificially prepared urine samples containing glucose, ketone, and protein. Actual pathological specimens containing glucose, protein, and nitrite also showed that instrumental readings were better than visual. Results of repeated visual readings spread over three different color blocks for certain concentrations of glucose, ketone, and protein, whereas repeated instrumental… CONTINUE READING
Recent Discussions
This paper has been referenced on Twitter 1 time over the past 90 days. VIEW TWEETS

From This Paper

Topics from this paper.

Citations

Publications citing this paper.
Showing 1-10 of 14 extracted citations

Similar Papers