Safe performance of magnetic resonance imaging on a patient with an ICD.

@article{Wollmann2005SafePO,
  title={Safe performance of magnetic resonance imaging on a patient with an ICD.},
  author={Christian G Wollmann and Matthias Grude and Bernd Tombach and Harald Kugel and Walter Heindel and G{\"u}nter Breithardt and Dirk Boecker and Christian Vahlhaus},
  journal={Pacing and clinical electrophysiology : PACE},
  year={2005},
  volume={28 4},
  pages={339-42}
}
This is a report on a patient with an implanted cardioverter defibrillator (ICD) who intentionally underwent magnetic resonance imaging (MRI) of a malignant brain tumor. To avoid inadequate detection of ventricular tachycardia (VT) or ventricular fibrillation (VF), the ICD was inactivated by programming the VT-detection and VT/VF-therapy status off. The patient came through the protocol safely and without any difficulty or discomfort. There was no arrhythmic event. MRI affected neither… CONTINUE READING