SU-F-T-138: Commissioning and Evaluating Dose Computation Models for a Dedicated Proton Line Scanning Beam Nozzle in Eclipse Treatment Planning System.

@article{Tsai2016SUFT138CA,
  title={SU-F-T-138: Commissioning and Evaluating Dose Computation Models for a Dedicated Proton Line Scanning Beam Nozzle in Eclipse Treatment Planning System.},
  author={P Tsai and H Huang and Shengbao Cai and Hsin-Yi Chen and S Wu and Tony Wu and Shwu-Hua Lee and Chien Yuh Yeh and Chung-Chi Lee},
  journal={Medical physics},
  year={2016},
  volume={43 6},
  pages={3493}
}
PURPOSE In this study, we present an effective method to derive low dose envelope of the proton in-air spot fluence at beam positions other than the isocenter to reduce amount of measurements required for planning commission. Also, we demonstrate commissioning and validation results of this method to the Eclipse treatment planning system (version 13.0.29… CONTINUE READING