SU-E-T-285: Flatness as a Measure of Changes in Photon Energy for Megavoltage X- Ray Radiotherapy.

@article{Gao2012SUET285FA,
  title={SU-E-T-285: Flatness as a Measure of Changes in Photon Energy for Megavoltage X- Ray Radiotherapy.},
  author={S Gao and Mark Rose and William E. Simon and Peter A. Balter},
  journal={Medical physics},
  year={2012},
  volume={39 6Part14},
  pages={
          3769
        }
}
PURPOSE To compare the sensitivity of flatness based energy metrics versus an attenuation based metric (percent depth dose) as a function of energy changes from the linac's nominal clinical energy. METHODS Energy changes were accomplished by adjusting the bending magnet current (BMI) ±15% away from the value used clinically. Two metrics for flatness, relative flatness in the central 80% of the field (flatness) and average maximum dose along the diagonals normalized by central axis (CAX) dose… 