SU-E-J-126: Respiratory Gating Quality Assurance: A Simple Method to Achieve Millisecond Temporal Resolution.

@article{McCabe2014SUEJ126RG,
  title={SU-E-J-126: Respiratory Gating Quality Assurance: A Simple Method to Achieve Millisecond Temporal Resolution.},
  author={Bradley P McCabe and Rodney D. Wiersma},
  journal={Medical physics},
  year={2014},
  volume={41 6},
  pages={185}
}
PURPOSE Low temporal latency between a gating on/off signal and a linac beam on/off during respiratory gating is critical for patient safety. Although, a measurement of temporal lag is recommended by AAPM Task Group 142 for commissioning and annual quality assurance, there currently exists no published method. Here we describe a simple, inexpensive, and… CONTINUE READING