Results of an international comparison for the activity measurement of 177Lu.

Abstract

An international Key Comparison of (177)Lu has recently been carried out. Twelve laboratories performed assays for radioactivity content on aliquots of a common master solution of (177)Lu, leading to eleven results submitted for entry into the Key Comparison Database of the Mutual Recognition Arrangement. A proposed Comparison Reference Value (CRV) was calculated to be 3.288(4)MBq/g using all eleven results. Degrees of equivalence and their uncertainties were calculated for each laboratory based on the CRV. Most of the values reported by the participating laboratories were within 0.6% of the CRV.

DOI: 10.1016/j.apradiso.2012.02.014

Cite this paper

@article{Zimmerman2012ResultsOA, title={Results of an international comparison for the activity measurement of 177Lu.}, author={B E Zimmerman and T Altzitzoglou and A Antohe and A Arinc and E Bakhshandeiar and D E Bergeron and L Bignell and C Bobin and M Capogni and J T Cessna and M L Cozzella and C J da Silva and P De Felice and M S Dias and T Dziel and A Fazio and R Fitzgerald and A Iwahara and F Jaubert and L Johansson and J Keightley and M F Koskinas and K Kossert and J Lubbe and A Luca and L Mo and O N{\"a}hle and O Ott and J Paepen and S Pomm{\'e} and M Sahagia and B R S Simpson and F F V Silva and R van Ammel and M J van Staden and W M van Wyngaardt and I M Yamazaki}, journal={Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine}, year={2012}, volume={70 9}, pages={1825-30} }