Residual ethylene oxide in hollow-fiber dialyzers.

@article{Henne1984ResidualEO,
  title={Residual ethylene oxide in hollow-fiber dialyzers.},
  author={W. Henne and W. Dietrich and M. Pelger and G. von Sengbusch},
  journal={Artificial organs},
  year={1984},
  volume={8 3},
  pages={
          306-10
        }
}
  • W. Henne, W. Dietrich, +1 author G. von Sengbusch
  • Published 1984
  • Medicine
  • Artificial organs
  • The deaeration and extraction kinetics of ethylene oxide (ETO) in hollow-fiber dialyzers were examined. The investigations showed that both deaeration and extraction can be described by two additive first-order kinetic expressions. The ETO content of membranes decreased rapidly at first, followed by a second slower decrease. Even after prolonged storage time, residual ETO was still detectable in the milligram range. These residuals cannot be eliminated by the usual rinsing of the dialyzer prior… CONTINUE READING
    37 Citations

    Topics from this paper

    A Case for Cellulosic Membrane Hemodialyzers
    • 1
    Biocompatibility of haemodialysis membranes.
    • 3
    Severe anaphylactoid reactions to cuprammonium cellulose hemodialyzers.
    • 60

    References

    Allergy to a product(s) of ethylene oxide gas: demonstration of IgE and IgG antibodies and hapten specificity.
    • 81