Refinements in the automated fluorometric histamine analysis system.

@article{Siraganian1975RefinementsIT,
  title={Refinements in the automated fluorometric histamine analysis system.},
  author={Reuben P. Siraganian},
  journal={Journal of immunological methods},
  year={1975},
  volume={7 2-3},
  pages={283-90}
}
The automated continuous flow system for the extraction and fluorometric analysis of histamine has been further improved. Samples with or without protein can be analyzed. The use of a single pump, added nitrogen, and lower concentration of o-phthalaldehyde increase the sensitivity by 5-10 fold and allow the analysis of samples at the rate of 30 per hour. Samples which are less than 0.5 ml in volume and contain 0.1 to 10 ng/ml histamine can routinely be analyzed. The sensitivity of the system is… CONTINUE READING

From This Paper

Topics from this paper.

Citations

Publications citing this paper.
Showing 1-10 of 24 extracted citations

Similar Papers

Loading similar papers…