Rate of wear in total hip replacement.

Abstract

The wear of high molecular weight polyethylene sockets was measured on radiographs, over a period of 9 to 10 years, in a group of patients whose ages averaged 73.3 years at the end of the period. The average rate of wear was 0.15 mm/year and 68 per cent of patients followed this pattern. Wear more than 2.5 mm in 10 years occurred in 15 per cent. There was a… (More)