Radiochromic film dosimetry in water phantoms.

@article{Butson2001RadiochromicFD,
  title={Radiochromic film dosimetry in water phantoms.},
  author={Martin J. Butson and Tsang Cheung and Peter K. N. Yu},
  journal={Physics in medicine and biology},
  year={2001},
  volume={46 1},
  pages={N27-31}
}
Radiochromic film is investigated for use in dosimetry in water phantoms as opposed to solid phantoms. Investigations are performed to measure the penetration rates of water into radiochromic film and to assess the effects on optical density that this penetration causes. The effects of film orientation during irradiation in water are also tested. Results show that only a small penetration rate is seen from water into the film which only affects the outer areas of the film, with penetration… CONTINUE READING

From This Paper

Figures, tables, and topics from this paper.

Citations

Publications citing this paper.
Showing 1-10 of 30 extracted citations

Is It Possible to Publish a Calibration Function for Radiochromic Film?

International journal of medical physics, clinical engineering and radiation oncology • 2014

References

Publications referenced by this paper.
Showing 1-7 of 7 references

Post irradiation dependence of radiochromic film optical density on temperature and time

L Reinstein, G Gluckman, S Pai, A Meek
Med. Phys • 1997
View 1 Excerpt

Post irradiation dependence of radiochromic film optical density on temperature and time Med

L Reinstein, G Gluckman, S Pai, A Meek
1997
View 1 Excerpt

Similar Papers