Quantitative analysis of small molecules by desorption electrospray ionization mass spectrometry from polytetrafluoroethylene surfaces.

Abstract

Desorption electrospray ionization mass spectrometry (DESI-MS) is an emerging technique for ambient analysis. However, its application to routine quantitation has not been explored extensively and this is undertaken here. We present studies that utilize a particularly suitable surface, porous polytetrafluoroethylene (porous PTFE), which shows less cross contamination between samples and improved sensitivity and signal stability compared to other surfaces. Quantitative experiments for 1 microL spotted solutions of the beta-blocker propranolol, using isotopically labeled propranolol as internal standard, showed a good linear correlation (r2 > 0.996) over the range 0.01-100 microM. The inter-day precision, based on the relative standard deviation, for replicates analyzed on three different days was 13% for 0.01 microM and better than 7% for the remainder of the calibration points. The inter-day accuracy, expressed as relative error, was better than +/-7% for all calibration points along the curve. These day-to-day measurements suggest that DESI-MS can be successfully employed for routine quantitative analysis. The use of the analog atenolol as an internal standard and further considerations that should improve quantitation by DESI-MS are also presented.

DOI: 10.1002/rcm.3377

Statistics

020402008200920102011201220132014201520162017
Citations per Year

60 Citations

Semantic Scholar estimates that this publication has 60 citations based on the available data.

See our FAQ for additional information.

Cite this paper

@article{Ifa2008QuantitativeAO, title={Quantitative analysis of small molecules by desorption electrospray ionization mass spectrometry from polytetrafluoroethylene surfaces.}, author={Demian R Ifa and Nicholas E Manicke and Allison L Rusine and R. Cooks}, journal={Rapid communications in mass spectrometry : RCM}, year={2008}, volume={22 4}, pages={503-10} }