Quantitative 3-dimensional echocardiography for accurate and rapid cardiac phenotype characterization in mice.

Abstract

BACKGROUND Insufficient techniques exist for rapid and reliable phenotype characterization of genetically manipulated mouse models of cardiac dysfunction. We developed a new, robust, 3-dimensional echocardiography (3D-echo) technique and hypothesized that this 3D-echo technique is as accurate as magnetic resonance imaging (MRI) and histology for assessment of left ventricular (LV) volume, ejection fraction, mass, and infarct size in normal and chronically infarcted mice. METHODS AND RESULTS Using a high-frequency, 7/15-MHz, linear-array ultrasound transducer, we acquired ECG and respiratory-gated, 500-microm consecutive short-axis slices of the murine heart within 4 minutes. The short-axis movies were reassembled off-line in a 3D matrix by using the measured platform locations to position each slice in 3D. Epicardial and endocardial heart contours were manually traced, and a B-spline surface was fitted to the delineated image curves to reconstruct the heart volumes. Excellent correlations were obtained between 3D-echo and MRI for LV end-systolic volumes (r=0.99, P<0.0001), LV end-diastolic volumes (r=0.99, P<0.0001), ejection fraction (r=0.99, P<0.0001), LV mass (r=0.94, P<0.0019), and infarct size (r=0.98, P<0.0001). Also, excellent correlations were found between the 3D-echo-derived LV mass and necropsy LV mass in normal mice (r=0.99, P<0.0001), as well as for 3D-echo-derived infarct size and histologically determined infarct size (r=0.99, P<0.0001) in mice with chronic heart failure. Bland-Altman analysis showed excellent limits of agreement between techniques for all measured parameters. CONCLUSIONS This new, fast, and highly reproducible 3D-echo technique should be of widespread applicability for high-throughput murine cardiac phenotyping studies.

7 Figures and Tables

02040'04'05'06'07'08'09'10'11'12'13'14'15'16'17
Citations per Year

117 Citations

Semantic Scholar estimates that this publication has 117 citations based on the available data.

See our FAQ for additional information.

Cite this paper

@article{Dawson2004Quantitative3E, title={Quantitative 3-dimensional echocardiography for accurate and rapid cardiac phenotype characterization in mice.}, author={Dana K. Dawson and Craig A. Lygate and James C. Saunders and J{\"{u}rgen E. Schneider and Xujiong Ye and Karen J. Hulbert and J. Alison Noble and Stefan Neubauer}, journal={Circulation}, year={2004}, volume={110 12}, pages={1632-7} }