Quantifying metal artefact reduction using virtual monochromatic dual-layer detector spectral CT imaging in unilateral and bilateral total hip prostheses.

@article{Wellenberg2017QuantifyingMA,
  title={Quantifying metal artefact reduction using virtual monochromatic dual-layer detector spectral CT imaging in unilateral and bilateral total hip prostheses.},
  author={Ruud H. H. Wellenberg and Martijn F. Boomsma and J. A. C. van Osch and Alain Vlassenbroek and Julien Milles and Mireille A. Edens and Geert J. Streekstra and C H Slump and Mario Maas},
  journal={European journal of radiology},
  year={2017},
  volume={88},
  pages={61-70}
}
PURPOSE To quantify the impact of prosthesis material and design on the reduction of metal artefacts in total hip arthroplasties using virtual monochromatic dual-layer detector Spectral CT imaging. METHODS The water-filled total hip arthroplasty phantom was scanned on a novel 128-slice Philips IQon dual-layer detector Spectral CT scanner at 120-kVp and 140-kVp at a standard computed tomography dose index of 20.0mGy. Several unilateral and bilateral hip prostheses consisting of different metal… CONTINUE READING