Quality assurance for image-guided radiation therapy utilizing CT-based technologies: a report of the AAPM TG-179.
@article{Bissonnette2012QualityAF, title={Quality assurance for image-guided radiation therapy utilizing CT-based technologies: a report of the AAPM TG-179.}, author={Jean-Pierre Bissonnette and Peter A. Balter and Lei Dong and Katja Langen and Dale Michael Lovelock and Moyed Miften and Douglas J. Moseley and Jean Pouliot and Jan-Jakob Sonke and Sua Yoo}, journal={Medical physics}, year={2012}, volume={39 4}, pages={ 1946-63 } }
PURPOSE
Commercial CT-based image-guided radiotherapy (IGRT) systems allow widespread management of geometric variations in patient setup and internal organ motion. This document provides consensus recommendations for quality assurance protocols that ensure patient safety and patient treatment fidelity for such systems.
METHODS
The AAPM TG-179 reviews clinical implementation and quality assurance aspects for commercially available CT-based IGRT, each with their unique capabilities and…
214 Citations
Image guidance doses delivered during radiotherapy: Quantification, management, and reduction: Report of the AAPM Therapy Physics Committee Task Group 180.
- Medicine, PhysicsMedical physics
- 2018
Balancing ALARA with the requirement for effective target localization requires that imaging dose be managed based on the consideration of weighing risks and benefits to the patient.
Physics support of image-guided radiotherapy.
- Medicine, PhysicsClinical oncology (Royal College of Radiologists (Great Britain))
- 2013
Assessment of interfraction patient setup for head-and-neck cancer intensity modulated radiation therapy using multiple computed tomography-based image guidance.
- Medicine, PhysicsInternational journal of radiation oncology, biology, physics
- 2013
Low cost multifunctional 3D printed image quality and dose verification phantom for an image-guided radiotherapy system
- Physics, MedicinePloS one
- 2022
The multifunctional phantom module meets the ACR recommended QA guidelines and is suitable for both LINACs and CT-sim, and meet the American College of Radiology (ACR) standards for QA of CT simulators and LINacs.
Rigid and Deformable Image Registration for Radiation Therapy: A Self-Study Evaluation Guide in YYYY Clinical Trial Participation.
- Medicine, PhysicsPractical radiation oncology
- 2021
Image quality and dose of an accelerator-integrated kV CBCT systems
- Medicine, Physics
- 2014
The research described in this dissertation was focused on investigation of the image quality of the planar and cone-beam computed tomography images of two imaging systems commonly used in radiotherapy: Varian On-Board Imager (OBI) and Elekta X-ray Volumetric Imagers (XVI).
Recommendations on the Practice of Calibration, Dosimetry, and Quality Assurance for Gamma Stereotactic Radiosurgery: Report of AAPM Task Group 178.
- Medicine, PhysicsMedical physics
- 2021
The TG-178 report recommends a GSR dose-rate calibration formalism and provides tabulated data to implement it for ionization chambers commonly used in GSR dosimetry and describes routine mechanical, dosimetric and safety checks for GSR devices.
Quality Assurance Results for a Commercial Radiosurgery System
- Medicine, PhysicsTechnology in cancer research & treatment
- 2015
With proper QA processes, the reported SRS system complies with tolerances set out in established guidelines, and discussions with the vendor are ongoing to address some of the hardware issues related to cone alignment.
The potential failure risk of the cone-beam computed tomography-based planning target volume margin definition for prostate image-guided radiotherapy based on a prospective single-institutional hybrid analysis
- Medicine, PhysicsRadiation oncology
- 2018
To determine a reliable CBCT-ST-based PTV margin, at least the systemic error Σ and the random error σ for on-line matching errors need to be investigated by supportive preliminary FM evaluation at least once.
Quality control in cone-beam computed tomography (CBCT) EFOMP-ESTRO-IAEA protocol (summary report).
- MedicinePhysica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics
- 2017
References
SHOWING 1-10 OF 228 REFERENCES
A quality assurance program for image quality of cone-beam CT guidance in radiation therapy.
- Physics, MedicineMedical physics
- 2008
The authors are presenting data demonstrating the high linearity of CT numbers, the uniformity of axial reconstructions, and the high contrast spatial resolution of ten CBCT systems from two commercial vendors, as well as demonstrating the long-term stability of the flat-panel detector and of the distances measured on reconstructed volumetric images.
Quality assurance for the geometric accuracy of cone-beam CT guidance in radiation therapy.
- Medicine, PhysicsInternational journal of radiation oncology, biology, physics
- 2008
Commissioning and clinical implementation of a mega-voltage cone beam CT system for treatment localization.
- Medicine, PhysicsMedical physics
- 2007
The commissioning and clinical implementation of the Siemens MVision mega-voltage cone beam CT (MV-CBCT) system is presented, the geometry and gain calibration procedures for the system are described, and guidelines for quality assurance procedures are provided.
Performance characteristics and quality assurance aspects of kilovoltage cone-beam CT on medical linear accelerator.
- Medicine, PhysicsMedical dosimetry : official journal of the American Association of Medical Dosimetrists
- 2007
Dose recalculation and the Dose-Guided Radiation Therapy (DGRT) process using megavoltage cone-beam CT.
- Medicine, PhysicsInternational journal of radiation oncology, biology, physics
- 2009
Quality assurance of image-guidance technologies.
- Medicine, PhysicsSeminars in radiation oncology
- 2007
Quality assurance of onboard megavoltage computed tomography imaging and target localization systems for on- and off-line image-guided radiotherapy.
- Medicine, PhysicsInternational journal of radiation oncology, biology, physics
- 2008
A study on adaptive IMRT treatment planning using kV cone-beam CT.
- Medicine, PhysicsRadiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology
- 2007
Evaluation of a contour-alignment technique for CT-guided prostate radiotherapy: an intra- and interobserver study.
- Medicine, PhysicsInternational journal of radiation oncology, biology, physics
- 2004
QA for helical tomotherapy: report of the AAPM Task Group 148.
- MedicineMedical physics
- 2010
The emphasis of the report is to describe the rationale for the proposed QA program and to provide example tests that can be performed, drawing from the collective experience of the task group members and the published literature.