Quality assessment of two lactate test strip methods suitable for obstetric use.

@article{Nordstrm1998QualityAO,
  title={Quality assessment of two lactate test strip methods suitable for obstetric use.},
  author={Lennart Nordstr{\"o}m and Selina Chua and Atish Roy and Sabaratnam Arulkumaran},
  journal={Journal of perinatal medicine},
  year={1998},
  volume={26 2},
  pages={83-8}
}
Accuracy of lactate determinations in cord blood was tested for one reflectometric (Accusport) and one amperometric (Lactate Pro) microvolume test strip lactate meter. Both meters, using a whole blood sample, measure lower levels of lactate than a reflectometric device considered as a reference method, which analyses lactate in plasma. Readings were unaffected irrespective of lactate concentrations for the Lactate Pro, whilst the Accusport overestimated low lactate concentrations and… CONTINUE READING