QDR 4500A DXA overestimates fat-free mass compared with criterion methods.

@article{Tylavsky2003QDR4D,
  title={QDR 4500A DXA overestimates fat-free mass compared with criterion methods.},
  author={Fran Tylavsky and Timothy G. Lohman and Barbara A. Blunt and Dale A. Schoeller and Thomas Fuerst and Jane A. Cauley and Michael C. Nevitt and Marjolein Visser and Tamara B. Harris},
  journal={Journal of applied physiology},
  year={2003},
  volume={94 3},
  pages={959-65}
}
This study evaluated the accuracy with which the dual-energy X-ray absorptiometer (Hologic QDR 4500A) measured fat-free mass (FFM), fat mass (FM), and hydration of FFM. In a study of 58 men and women (ages 70-79 yr), the QDR 4500A was found to provide a systematically higher estimate of FFM and lower estimate of FM than a four-component model of body composition. A correction factor from this study was developed and applied to two other samples (n = 13 and 37). We found mean corrected levels of… CONTINUE READING