Proton beam therapy interference with implanted cardiac pacemakers.

@article{Oshiro2008ProtonBT,
  title={Proton beam therapy interference with implanted cardiac pacemakers.},
  author={Y. Oshiro and S. Sugahara and M. Noma and Masato Sato and Y. Sakakibara and T. Sakae and Yasutaka Hayashi and H. Nakayama and K. Tsuboi and N. Fukumitsu and Ayae Kanemoto and T. Hashimoto and K. Tokuuye},
  journal={International journal of radiation oncology, biology, physics},
  year={2008},
  volume={72 3},
  pages={
          723-7
        }
}
PURPOSE To investigate the effect of proton beam therapy (PBT) on implanted cardiac pacemaker function. METHODS AND MATERIALS After a phantom study confirmed the safety of PBT in patients with cardiac pacemakers, we treated 8 patients with implanted pacemakers using PBT to a total tumor dose of 33-77 gray equivalents (GyE) in dose fractions of 2.2-6.6 GyE. The combined total number of PBT sessions was 127. Although all pulse generators remained outside the treatment field, 4 patients had… Expand
Malfunctions of implantable cardiac devices in patients receiving proton beam therapy: incidence and predictors.
Radiation therapy with implanted cardiac pacemaker devices: a clinical and dosimetric analysis of patients and proposed precautions.
Case study thoracic radiotherapy in an elderly patient with pacemaker: the issue of pacing leads.
Influence of secondary neutrons induced by proton radiotherapy for cancer patients with implantable cardioverter defibrillators
Malfunctions of implantable cardiac devices in patients receiving proton beam therapy: incidence and predictions. In regard to Gomez et al.
  • D. Mihailidis
  • Medicine
  • International journal of radiation oncology, biology, physics
  • 2014
Epithermal neutron beam interference with cardiac pacemakers.
  • H. Koivunoro, T. Serén, +8 authors S. Savolainen
  • Chemistry, Medicine
  • Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine
  • 2011
Radiotherapy in patients with pacemakers and implantable cardioverter defibrillators: a literature review.
  • T. Zaremba, A. Jakobsen, M. Søgaard, A. Thøgersen, S. Riahi
  • Medicine
  • Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology
  • 2016
...
1
2
3
4
...

References

SHOWING 1-10 OF 17 REFERENCES
Radiotherapy to patients with artificial cardiac pacemakers.
Pacemaker failure associated with therapeutic radiation.
Pacemaker Failure Due to Radiation Therapy