Precursor and Neutral Loss Scans in an RF Scanning Linear Quadrupole Ion Trap

@article{Snyder2018PrecursorAN,
  title={Precursor and Neutral Loss Scans in an RF Scanning Linear Quadrupole Ion Trap},
  author={Dalton T Snyder and Lucas Szalwinski and Robert L. Schrader and V Pirro and Ryan T. Hilger and R. Graham Cooks},
  journal={Journal of The American Society for Mass Spectrometry},
  year={2018},
  volume={29},
  pages={1345-1354}
}
AbstractMethodology for performing precursor and neutral loss scans in an RF scanning linear quadrupole ion trap is described and compared to the unconventional ac frequency scan technique. In the RF scanning variant, precursor ions are mass selectively excited by a fixed frequency resonance excitation signal at low Mathieu q while the RF amplitude is ramped linearly to pass ions through the point of excitation such that the excited ion’s m/z varies linearly with time. Ironically, a nonlinear… CONTINUE READING