Photocurrent Measurement of PC and PV HgCdTe Detectors

@inproceedings{Eppeldauer2001PhotocurrentMO,
  title={Photocurrent Measurement of PC and PV HgCdTe Detectors},
  author={George P. Eppeldauer and Robert J. Martin},
  booktitle={Journal of research of the National Institute of Standards and Technology},
  year={2001}
}
Novel preamplifiers for working standard photoconductive (PC) and photovoltaic (PV) HgCdTe detectors have been developed to maintain the spectral responsivity scale of the National Institute of Standards and Technology (NIST) in the wavelength range of 5 μm to 20 μm. The linear PC mode preamplifier does not need any compensating source to zero the effect of the detector bias current for the preamplifier output. The impedance multiplication concept with a positive feedback buffer amplifier was… CONTINUE READING

References

Publications referenced by this paper.
Showing 1-6 of 6 references

Electric Circuits

  • J. W. Nilsson
  • 3rd Edition, Addison Wesley Publishing Company…
  • 2001
Highly Influential
4 Excerpts

Linear HgCdTe radiometer

  • G. Eppeldauer, L. Novak
  • SPIE Proc. Vol. 1110, A. J. Huber, M. J. Triplett…
  • 1989
Highly Influential
4 Excerpts

Noise characteristics of bootstrapped photovoltaic and photoconductive detectors

  • J. P. Makai, J. J. Makai, J. Balazs
  • SPIE Proc., S. M. Goodnick, W. F. Kailey, R. E…
  • 1999
2 Excerpts

Spectroradiometric Detector Measurements: Part III—Infrared Detectors

  • A. L. Migdall, G. P. Eppeldauer
  • NIST Special Publication 250-42
  • 1998
1 Excerpt

Photodiode Amplifiers

  • J. Graeme
  • McGraw-Hill, New York
  • 1996
2 Excerpts

Similar Papers