Performance of the compensation comparison method for retinal straylight measurement: effect of patient's age on repeatability.

Abstract

AIM The assessment of repeatability and reproducibility of retinal straylight measurements with the C-Quant straylight meter (Oculus AG, Germany) and the effect of patient's age on the instrument performance are tested with a series of experiments. METHODS First, 20 eyes from 20 subjects (mean age 26.9 (SD 2.7) years, mean refractive error -1.34 (2.72) D… (More)
DOI: 10.1136/bjo.2007.131078

6 Figures and Tables

Topics

  • Presentations referencing similar topics