Performance evaluation of a CyberKnife G4 image-guided robotic stereotactic radiosurgery system.

@article{Antypas2008PerformanceEO,
  title={Performance evaluation of a CyberKnife G4 image-guided robotic stereotactic radiosurgery system.},
  author={Christos E Antypas and Evaggelos Pantelis},
  journal={Physics in medicine and biology},
  year={2008},
  volume={53 17},
  pages={4697-718}
}
The aim of the current work was to present the performance evaluation procedures implemented at our department for the commissioning of a G4 CyberKnife system. This system consists of a robotic manipulator, a target-locating system and a lightweight 6-MV linac. Individual quality assurance procedures were performed for each of the CyberKnife subsystems. The system was checked for the mechanical accuracy of its robotic manipulator. The performance of the target-locating system was evaluated in… CONTINUE READING

Citations

Publications citing this paper.
Showing 1-10 of 36 extracted citations

References

Publications referenced by this paper.
Showing 1-10 of 52 references

2004a Measurements of the relative output factors for CyberKnife collimators Neurosurgery

  • C Yu, G Jozsef, L JApuzzoM, Z Petrovich
  • 2004
Highly Influential
7 Excerpts

Absorbed dose determination in external beam radiotherapy: an international code of practice for dosimetry based on standards of absorbed dose to water IAEA Technical Report Series 398 (Vienna: IAEA

  • P Andreo, D TBurns, +5 authors S Vynckier
  • 2000
Highly Influential
3 Excerpts

A comparison of dosimetry techniques in stereotactic radiosurgery

  • M Heydarian, P WHoban, A HBeddoe
  • Phys. Med. Biol
  • 1996
Highly Influential
5 Excerpts

A comparison of dosimetry techniques in stereotactic radiosurgery Phys

  • M Heydarian, W HobanP, H BeddoeA
  • . Med . Biol .
  • 1996
Highly Influential
5 Excerpts

Similar Papers