Performance characteristics of dedicated molecular breast imaging systems at low doses.

@article{Long2016PerformanceCO,
  title={Performance characteristics of dedicated molecular breast imaging systems at low doses.},
  author={Zaiyang Long and Amy Lynn Conners and Katie N. Hunt and Carrie B. Hruska and Michael K. O'Connor},
  journal={Medical physics},
  year={2016},
  volume={43 6},
  pages={3062-3070}
}
PURPOSE The purpose of this study was to compare the system performance characteristics and lesion detection capability of two molecular breast imaging (MBI) systems: a multicrystal sodium iodide (NaI)-based single-head system and a cadmium zinc telluride (CZT)-based dual-head system at low administered doses (150-300 MBq) of Tc-99m sestamibi. METHODS System performance characteristics including count sensitivity, uniformity, energy resolution, and spatial resolution were measured using… CONTINUE READING