OrthoPilot total hip arthroplasty workflow and surgery.

@article{Kiefer2005OrthoPilotTH,
  title={OrthoPilot total hip arthroplasty workflow and surgery.},
  author={Hartmuth Kiefer and Aneis Othman},
  journal={Orthopedics},
  year={2005},
  volume={28 10 Suppl},
  pages={
          s1221-6
        }
}
To analyze the workflow and navigation principles of the OrthoPilot total hip arthroplasty (THA) Version 2.0 (B. Braun-Aesculap, Tuttlingen, Germany), a consecutive series of 60 patients was used to compare the navigated intraoperative data and postoperative measurements of stem and cup position. Within the safe zone, 96.3% of 54 acetabular cups were positioned. The rotational accuracy of femoral instruments was 65 degrees in 41 THAs. The femoral offset was medialized by 6.05 mm in 76% and… CONTINUE READING
BETA

Similar Papers

Topics from this paper.

Citations

Publications citing this paper.
SHOWING 1-10 OF 14 CITATIONS

Design of a computer-aided visual system for Total Hip Replacement surgery

  • 2015 IEEE International Symposium on Circuits and Systems (ISCAS)
  • 2015
VIEW 1 EXCERPT
CITES METHODS

Orientation and depth estimation for femoral components using image sensor, magnetometer and inertial sensors in THR surgeries

  • 2015 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)
  • 2015