Optimization of the imaging protocol of an X-ray CT scanner for evaluation of normoxic polymer gel dosimeters

@inproceedings{Subramanian2006OptimizationOT,
  title={Optimization of the imaging protocol of an X-ray CT scanner for evaluation of normoxic polymer gel dosimeters},
  author={Brindha Subramanian and Paul B Ravindran and Clive Baldock},
  booktitle={Journal of medical physics},
  year={2006}
}
X-ray computer tomography (CT) has previously been reported as an evaluation tool for polymer gel (PAG) dosimeters. In this study, the imaging protocol of a Siemens Emotion X-ray CT scanner was optimized to evaluate PAGAT normoxic gel dosimeters. The scan parameters were optimized as 130 kV and 150 mA with a slice thickness of 3 mm for smaller fields and 5 mm for larger fields of irradiation. The number of images to be averaged to reduce noise to an acceptable level was concluded to be 25. It… CONTINUE READING