Optimization of Data Coils in a Multiband Wireless Link for Neuroprosthetic Implantable Devices

@article{Jow2010OptimizationOD,
  title={Optimization of Data Coils in a Multiband Wireless Link for Neuroprosthetic Implantable Devices},
  author={Uei-Ming Jow and Maysam Ghovanloo},
  journal={IEEE Transactions on Biomedical Circuits and Systems},
  year={2010},
  volume={4},
  pages={301-310}
}
We have presented the design methodology along with detailed simulation and measurement results for optimizing a multiband transcutaneous wireless link for high-performance implantable neuroprosthetic devices. We have utilized three individual carrier signals and coil/antenna pairs for power transmission, forward data transmission from outside into the body, and back telemetry in the opposite direction. Power is transmitted at 13.56 MHz through a pair of printed spiral coils (PSCs) facing each… CONTINUE READING