Abstract

Notching of the anterior femoral cortex during primary total knee arthroplasty can occur when making an anterior femoral bone cut. To examine the increased stress caused by a notch, we used a validated 3-dimensional finite element model of the femur under gait loads. Three factors that affected the stress concentration were identified: First, larger notches associated with increased stress concentrations. Second, sharper notches led to increased local stresses. Third, the proximity of the notch to the prostheses affected the stress concentration. In summary, anterior femoral notches greater than 3 mm with sharp corners located directly at the proximal end of the prosthesis produced the highest stress concentrations and may lead to a significant risk of periprosthetic fracture.

050200920102011201220132014201520162017
Citations per Year

208 Citations

Semantic Scholar estimates that this publication has 208 citations based on the available data.

See our FAQ for additional information.

Cite this paper

@article{Zalzal2006NotchingOT, title={Notching of the anterior femoral cortex during total knee arthroplasty characteristics that increase local stresses.}, author={Paul Zalzal and David Backstein and Allan E. Gross and Marcello Papini}, journal={The Journal of arthroplasty}, year={2006}, volume={21 5}, pages={737-43} }